FibroTest and Fibroscan performances revisited in patients with chronic hepatitis C. Impact of the spectrum effect and the applicability rate.

Clinics and research in hepatology and gastroenterology

Poynard T, De Lédinghen V, Zarski JP, Stanciu C, Munteanu M, Vergniol J, France J, Trifan A, Moussalli J, Lebray P, Thabut D, Ratziu V

2011 Clin Res Hepatol Gastroenterol Volume 35 Issue 11

PubMed 21852224 DOI 10.1016/j.clinre.2011.07.003

FibroTest Reliability vs. Elastography HCV Fibrosis Blood donors

BACKGROUND

Two widely used biomarkers of fibrosis, FibroTest and liver stiffness measurement (LSM), have been mostly validated in patients with chronic hepatitis C (CHC) using the standard area under the ROC curve (sAUROC) which is not the most appropriate method due to the risk of fibrosis spectrum effect. Furthermore the performance of these biomarkers have not been assessed in "intention to diagnose" which takes into account the failures and non-reliable results.

AIM

The aim was to compare the accuracy of FibroTest and LSM for the diagnosis of fibrosis using sAUROC, the pairwise comparison of fibrosis stages by Obuchowski measure (wAUROC), and these AUROCs reassessed after taking into account the applicability rates.

METHODS

One thousand two hundred and eighty-nine patients with CHC and 604 healthy volunteers were analyzed. The performances of biomarkers assessed were compared in a patients-only group (P1: n=1289), in a population combining both patients and healthy volunteers (P2: n=1893) and in a simulated population (P3: n=1893) with the prevalence of stages observed in a reference population, to demonstrate the impact of spectrum effect. Applicability rates were estimated prospectively in 24,872 consecutive FibroTest and in 13,669 consecutive LSM examinations.

RESULTS

Using wAUROC, the conclusions of studies with reliable results in P1 were different than in those of P2 and in P3. There was a lower performance of FibroTest versus LSM in P1 (0.864 [0.855-0.873] vs. 0.883 [0.874-0.892]; P=0.002) which was not found in P2 (0.893 [0.887-0.900] vs. 0.894 [0.887-0.901]; P=0.86) and in P3 (0.899 [0.893-0.905] vs 0.902 [0.895-0.909]; P=0.60). Using the sAUROC, in P1, P2 and P3, there was no significant difference between FibroTest and LSM performance for advanced fibrosis and a lower performance of FibroTest versus LSM for cirrhosis. In intention to diagnose, using wAUROCs performances were higher for FibroTest vs. LSM in P1 (0.857 [0.848-0.866] vs. 0.814 [0.807-0.821]; P<0.0001) and P2 (0.885 [0.879-0.892] vs. 0.743 [0.737-0.749]; P<0.0001), without difference in P3 (0.891 [0.885-0.897] vs. 0.894 [0.887-0.901]; P=0.90). Using sAUROC, the significant differences in favor of FibroTest vs LSM persisted also for the diagnosis of advanced fibrosis, both in P1 and P2 (P<0.0001) and for the diagnosis of cirrhosis in P1 (P<0.001).

CONCLUSION

When the spectrum effects and applicability rates were taken into account, LSM had lower performance results than FibroTest for the diagnosis of fibrosis stages.


Citation Reference: